%0 Journal Article %A LI Nan %A LI Xiu-ping %A REN Zheng %T Calibration Standard Design Based on the Microstrip Dual-Port Test %D 2013 %R 10.13190/j.jbupt.2013.05.001 %J Journal of Beijing University of Posts and Telecommunications %P 1-5 %V 36 %N 5 %X

Short、open、load、through (SOLT) and through、reflect、 line (TRL) calibration techniques were described. The 12-term system errors for two-port calibration were shown, and the actual S-parameters device was derived from the measured S-parameters by using matlab programs. Comparison between the matlab programs results and the vector network analyzer measurement data validated the correctness of the calibration approaches and procedures. The SOLT and TRL microstrip calibration standards were designed for microstrip filter measurement.

%U https://journal.bupt.edu.cn/EN/10.13190/j.jbupt.2013.05.001